Instrumento Aire

Volver

Powerex instrument air systems meet NFPA 99 requirements and are designed to provide clean, oil-less air to hospitals and medical institutions.

Solicitar un presupuesto
ModeloSistema HPBomba HPSCFM @ 185 PSIGDepósito (gal)dB(A) NivelFLA @ 208V FLA @ 230VFLA @ 460VHoja de especificacionesFolleto
Sistemas dúplex
IOPD07547.5 (2)7.520.41207643.039.020.5
IOPD100410 (2)1026.01207656.851.626.8
IOPD150415 (2)1534.31207984.276.439.2
Sistemas Triplex
IOPT0754 7.5 (3)7.540.81207963.557.529.75
IOPT100510 (3)1052.02007984.276.439.2
IOPT105615 (3)1568.624082125.3113.657.8
Sistemas Quadplex
IOPDQ07557.5 (4)7.561.22008184.076.039.0
IOPDQ00610 (4)1078.024081111.6101.251.6
IOPDQ150615 (4)15102.924084166.4150.876.4